Maximum 9 waveforms
Use built-in system, high integrated intelligent module technology, full digital monitoring, super functions, lowest * swearing and low failure rate
Features:
- 12.1 inch super resolution color TFT display
- Maximum 9 waveforms
- Use built-in system, high integrated intelligent module technology, full digital monitoring, super functions, lowest * swearing and low failure rate
- Combined functions of maternal monitor and fetal monitor
- Measure the intrapartum patient and fetal heart rate
- Uterine contraction and fetal movement marker
- Continuous real time recording of fetal heart rate
- Can be switched to normal patient monitoring interface
- 72 hours graphic trends and tabular of all parameters
- 1440 groups of NIBP data storage
- ST analysis
- Anti-high frequency electronic surgical unit and anti-defibrillation, for monitoring in the surgery or for cardiac care
- Use wide-angle, multi-wafer, wide beam and ultrasonic probes
Standard configuration:
- FHR, Uterine Contraction Pressure(TOCO), FM and Maternal ECG, NIBP, SPO2, TEMP, RESP, PULSE, Built-in rechargeable battery, thermal printer
Optional:
- Dual TEMP
Technical Specifications:
|
Safety |
IEC60601 approved (GB9706) |
|
Power source |
AC100V-240V, 50/60Hz |
|
Size |
|
|
Net weight |
|
|
Working |
Temperature: 5 ~ |
|
Transportation and Storage |
Temperature: -20 ~ |
Ultrasonic Probe: (Fetal Heart Rate)
|
Type |
Type B proof, dirctional, using pulse doppler system |
|
Working frequency |
2.5 MHz |
|
Transmitting power density |
|
|
Probe |
Multi-wafer probe |
|
Sample rate |
5 ms |
|
Calculating precision of FHR |
0.25 bpm |
|
Accuracy of FHR |
1 bpm (100-180 bpm) |
Marker (Fetal Movement)
|
Manually push button mode |
|
When pressed, a red strip will be displayed above the FHR graph |
|
Uterine Contraction Pressure |
|
Type B proof |
|
Uterine contraction pressure probe |
|
Manual zero adjustment |
|
Design pressure sensitivity: 0-100% |
ECG
|
Lead |
5 leads (R, L, F, N, C) |
|
Lead selection |
I, II, III, avR, avL, avF, V |
|
Waveform |
7 channels |
|
Gain selection |
×0. 5, ×1, ×2 |
|
HR and alarm range |
Adult: 15-300bpm |
|
Accuracy |
±2% or ±2bpm, whichever is greater |
|
Input impedance |
> 5ΜΩ |
|
Voltage range of electrode polarization |
±300mV |
|
Leakage current |
< 30uA |
|
Filter |
Diagnostic mode: 0.05 ~ 100Hz |
|
Scaling signal |
1mV ± 3% |
|
Measurement range |
-1.0mV ~ +1.0mV |
|
|
-0.80mV~0.80mV |
RESP
|
Model |
RA-LL impedance |
|
Measurement range |
Adult: 0 ~ 120bpm |
|
Resolution |
1rpm |
|
Accuracy |
±2rpm |
NIBP
|
Method |
Automatic oscillometric |
|
Operation mode |
Auto/Manual |
|
Alarm Type |
Systolic, Diastolic, Mean |
|
Measurement ranges |
Systolic BP: 40 ~ 300mmHg |
|
Resolution |
1mmHg |
|
Accuracy |
Max average error ±5mmHg or ±5%, whichever is greater |
|
Overload Protection |
Double safety protection |
SPO2
|
|
0 ~ 100% |
|
Alarm range |
0 ~ 100% |
|
Resolution |
1% |
|
Accuracy |
±2% (80~100%) |
PULSE RATE
|
Range |
0 ~ 254bpm |
|
Resolution |
1bpm |
|
Accuracy |
±1bpm or ±2%, whichever is greater |
|
Alarm delay time |
10sec |
TEMP
|
Channel |
2 channels |
|
Measurement range |
0~ |
|
Alarm range |
0~ |
|
Resolution |
|
|
Accuracy |
± |
Tell us what you think about this item and share your opinions with other people. Please make sure your review focuses only on this item. All reviews are moderated and will be reviewed within two business days. Inappropriate reviews will not be posted.
After-sales questions and issues: Contact our Customer Service Department. (A customer representative will get back to you).
Please Sign In first if you want to later edit or manage reviews uploaded by yourself.



The safer, easier way to pay online
View demo
U.S.Dollar
Euro
U.K.Pound
Australian Dollar
CANADA Dollar

View larger image
Bookmark this product
Add to My Favorites